User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Fluke 702 Sub 3 DATE: 10-Oct-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 190 CONFIGURATION: Fluke 5700A CONFIGURATION: Datron 1281 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD {* T/C MEASURE TEST *} 1.004 HEAD { Type "K"} 1.005 HEAD * T/C MEASURE TEST * 1.006 JMP 2.001 1.007 EVAL 2.001 PIC 55_70xTC 2.002 DISP Insert the reference thermocouple into the center 2.002 DISP well of the Ice Point Cell. 2.003 DISP Connect the copper end (yellow dual banana connector) 2.003 DISP to the 5700 OUTPUT HI/LO. 2.004 DISP Connect a "K" type connector with "K" type T/C wire 2.004 DISP to UUT thermocouple input. 2.005 DISP Press the TC/RTD button to select TC Measure. 2.006 DISP Use the arrow buttons to select type K and then 2.006 DISP press ENTER. 2.007 JMP 3.001 2.008 EVAL 3.001 5700 -5.550mV S 2W 3.002 MEMI Use the keyboard to enter the UUT reading in °C. 3.003 ACC -180.0°C 0.10U 3.004 MEME 3.005 MEMC °C 1.0U #! Test Tol 1, Sys Tol 0.1, TUR 10.000 (>= 4.00). #52.005 5500 -180.0degC 1.0U _K TC 2W #53.001 5500 0.0degC 0.5U _K TC 2W 4.001 5700 0.000mV S 2W 4.002 MEMI Use the keyboard to enter the UUT reading in °C. 4.003 ACC 0.0°C 0.1U 4.004 MEME 4.005 MEMC °C 0.5U #! Test Tol 0.5, Sys Tol 0.1, TUR 5.000 (>= 4.00). #54.001 5500 1300.0degC 0.9U _K TC 2W 5.001 5700 52.410mV S 2W 5.002 MEMI Use the keyboard to enter the UUT reading in °C. 5.003 ACC 1300.0°C 0.1U 5.004 MEME 5.005 MEMC °C 0.9U #! Test Tol 0.9, Sys Tol 0.1, TUR 9.000 (>= 4.00). 6.001 ASK- N 6.002 DISP Disconnect the test setup. 6.003 HEAD {* T/C SOURCE TEST *} 6.004 JMP 7.001 6.005 EVAL 7.001 IEEE [@1281] DCV AUTO 7.002 DISP Connect the copper end of the "K" type reference probe 7.002 DISP to the Hi/Lo inputs on the 1281. 7.003 DISP Press MEAS/SOURCE to select Source. 7.004 DISP Press the TC/RTD button to select TC Source. 7.005 DISP Use the arrow buttons to select type K and then 7.005 DISP press ENTER. 7.006 DISP Press ENTER to remove Warning screen. 7.007 DISP Press -200 and then ENTER for -200.0 °C. 7.008 ACC 200 -5.891mV 7P% 0.5P/ 7.009 IEEE [@1281] RDG?[I] 7.010 MEM* 1000 7.011 MEME 7.012 MEMC mV 0.0066U -200°C #! Test Tol 0.016, Sys Tol 0.00014848, TUR 107.757 (>= 4.00). #55.008 5500 -200.0degC 1.0U _K TM 2W 8.001 DISP Press 0 and then ENTER for 0.0degC. #56.002 5500 0.0degC 0.5U _K TM 2W 8.002 ACC 200 0.000mV 0.5P/ 8.003 IEEE [@1281] RDG?[I] 8.004 MEM* 1000 8.005 MEME 8.006 MEMC mV 0.0117U 0°C #! Test Tol 0.02, Sys Tol 0.00010725, TUR 186.477 (>= 4.00). 9.001 DISP Press 1370 and then ENTER 1370.0degC. #57.002 5500 1370.0degC 0.9U _K TM 2W 9.002 ACC 200 54.819mV 7P% 0.5P/ 9.003 IEEE [@1281] RDG?[I] 9.004 MEM* 1000 9.005 MEME 9.006 MEMC mV 0.0099U 1370°C #! Test Tol 0.033, Sys Tol 0.00047654, TUR 69.249 (>= 4.00). 10.001 DISP Press CLEAR (ZERO) twice to turn off the output. 10.002 ASK+ N 10.003 HEAD {* mA MEASURE TEST *} 10.004 HEAD { 30mA Range} 10.005 HEAD * mA MEASURE TEST * 10.006 JMP 11.001 10.007 EVAL 11.001 PIC 55_70x_A 11.002 DISP Press MEAS/SOURCE twice to select Measure. 11.003 DISP Press mA to select DCA Measure. 11.004 MMFC C1 11.005 5700 30 4.000mA 0.009U 2W #! Test Tol 9e-006, Sys Tol 3e-007, TUR 30.000 (>= 4.00). 12.001 5700 30 20.000mA 0.013U 2W #! Test Tol 1.3e-005, Sys Tol 1.1e-006, TUR 11.818 (>= 4.00). 13.001 5700 30 30.000mA 0.015U 2W #! Test Tol 1.5e-005, Sys Tol 2.8e-006, TUR 5.357 (>= 4.00). 14.001 5700 30 -30.000mA 0.015U 2W #! Test Tol 1.5e-005, Sys Tol 2.8e-006, TUR 5.357 (>= 4.00). 15.001 HEAD mA MEASURE: {100mA Range} 15.002 JMP 16.001 15.003 EVAL 16.001 5700 100 0.00mA 0.06U 2W #! Test Tol 6e-005, Sys Tol 1e-008, TUR 6000.000 (>= 4.00). 17.001 5700 100 100.00mA 0.11U 2W #! Test Tol 0.00011, Sys Tol 7e-006, TUR 15.714 (>= 4.00). 18.001 5700 100 -100.00mA 0.11U 2W #! Test Tol 0.00011, Sys Tol 7e-006, TUR 15.714 (>= 4.00). 19.001 ASK- R N 19.002 HEAD {* SIMULATE TRANSMITTER TEST *} 19.003 JMP 20.001 19.004 EVAL 20.001 PIC 58_70xST 20.002 IEEE [@1281] DCI AUTO 20.003 DISP Press MEAS/SOURCE to select Source. 20.004 DISP Press mA. 20.005 DISP Use the arrow buttons to select Simulate Transmitter 20.005 DISP and then press ENTER. 20.006 DISP Press ENTER to remove Warning screen. 20.007 5700 4V S 2W 20.008 DISP Press 4 and then ENTER for 4.000mA. #15.014 3458 22 4.000mA 0.021U E 2W 20.009 IEEE [@1281] [D1000] RDG?[I] 20.010 MEM* 1000 20.011 ACC 20 4.000mA 100P% 20P/ 20.012 MEME 20.013 MEMC mA 0.021U #! Test Tol 0.021, Sys Tol 0.0008, TUR 26.250 (>= 4.00). 21.001 5700 4V S 2W 21.002 DISP Press 22 and then ENTER for 22.000mA. #17.003 3458 11 22.000mA 0.066U E 2W 21.003 IEEE [@1281] [D1000] RDG?[I] 21.004 MEM* 1000 21.005 ACC 200 22.000mA 100P% 20P/ 21.006 MEME 21.007 MEMC mA 0.066U #! Test Tol 0.066, Sys Tol 0.0062, TUR 10.645 (>= 4.00). 22.001 DISP Press CLEAR (ZERO) twice to turn off the output. 22.002 DISP Disconnect the test setup. 22.003 HEAD {* mA SOUCRE TEST *} 22.004 JMP 23.001 22.005 EVAL #17.008 PIC 58_70x_A #17.011 IEEE [@3458]DCI 23.001 DISP Connect the 1281 and the UUT as follows. 23.001 DISP 1281 TO 702 23.001 DISP AMP I+ ............... mA SOURCE HI 23.001 DISP AMPS I- ............... mA SOURCE LO 23.002 DISP Press the mA button. 23.003 DISP Use the arrow buttons to select Source mA and then 23.003 DISP press ENTER. 23.004 DISP Press 2 and then ENTER for 2.000mA. #17.013 3458 2.000mA 0.009U E 2W 23.005 IEEE [@1281] RDG?[I] 23.006 MEM* 1000 23.007 ACC 20 2.000mA 100P% 20P/ 23.008 MEME 23.009 MEMC mA 0.009U #! Test Tol 0.009, Sys Tol 0.0006, TUR 15.000 (>= 4.00). #18.001 IEEE [@3458]DCI 24.001 DISP Press 4 and then ENTER for 4.000mA. #18.003 3458 4.000mA 0.009U E 2W 24.002 IEEE [@1281] RDG?[I] 24.003 MEM* 1000 24.004 ACC 20 4.000mA 100P% 20P/ 24.005 MEME 24.006 MEMC mA 0.009U #! Test Tol 0.009, Sys Tol 0.0008, TUR 11.250 (>= 4.00). 25.001 DISP Press 22 and then ENTER for 22.000mA. #9.003 3458 22.000mA 0.011U E 2W 25.002 IEEE [@1281] RDG?[I] 25.003 MEM* 1000 25.004 ACC 200 22.000mA 100P% 20P/ 25.005 MEME 25.006 MEMC mA 0.011U #! WARNING: Test Tol 0.011, Sys Tol 0.0062, TUR 1.774 (< 4.00). 26.001 DISP Press CLEAR (ZERO) twice to turn off the output. 26.002 DISP Disconnect the test setup. 26.003 IEEE [@1281]*RST 26.004 END